Improving Autofocusing
Because MSI-raster is normally used for acquiring images on negatively stained particles on continuous carbon, the intermediate mag preset (hl) often lacks contrast. To reduce focusing failure, we recommend a different focusing sequence for the Z focus node.
-
Use Stage Alpha Tilt to determine defocus at the low magnification sq preset to do a rough stage-z adjustment to U-center
-
Refine stage-z position at the high magnification fa preset and Use beam-tilt to determine the required correction.
For "Stage_Wobble" Step:
-
Enable = Yes
-
Image registration= phase correlation
-
Fit limit = 1000 (not used)
-
Correct for delta Defocus/Z between 0 and 2e-4 meters
-
Correction type = Stage Z
-
Check drift greater than 3e-10 meters/s = No
-
Stigmator correction = No
-
Stigmator Defocus Max/Min = N/A
For "Z_to_Eucentric" Step, change these:
-
preset = fa
-
Correct for delta Defocus/Z between 0 and 2e-5 meters