Troubles with Calibrations
-
Available magnification from pixel size calibration is different from that said on the scope
-
Stage does not move after a pyScope update on the computer controlling the microscope
Magnification list not consistent with that at the microscope
Commonly Why: Instrument was not selected correctly when Leginon is installed
Solution: Check the parameters of the named instrument in Administration Tool
Export calibrations as json file and send to development team
Once you have a session that successfully use the settings that you would like to distribute
cd /your_myami/dbschema/tools
python ./export_leginon_settings.py source_database_hostname source_camera_hosthame camera_name
For example,
python ./export_leginon_settings.py my_leginon_dbhost krios-k2 GatanK2Counting